Sito
segnalato su: |
|
|
|
|
|
|
richiedi
un
preventivo
di
dosimetri
 |
|
|
The area monitors of the type "WS 05 C" are stationary room monitoring
systems for the dose rate measurement of gamma radiation and X-rays,
preferably used for the permanent monitoring of isotope laboratories,
radiotherapy rooms, storage rooms for radioactive materials, etc..
The "WS 05 C" is available in the following versions:
-
WS 05
C-1
Area monitor (1 measuring channel) with RS-232-interface for the
operation with any probe of the GRAETZ-programme.
-
WS 05
C-2
Area monitor (2 measuring channels) with RS-232-interface for one
measuring channel, for the simultaneous operation of max. 2
different or similar probes - any of the GRAETZ-programme.
-
WS 05
C-3 (see picture)
Area monitor (3 measuring channels) with RS-232-interface for one
measuring channel, for the simultaneous operation of max. 3
different or similar probes - any of the GRAETZ-programme.
Each measuring channel indicates the
measured values on a digital, illuminated LC-display (60x30 mm). For
each measuring channel four individual dose rate alarm thresholds are
free programmable over the whole measuring range of the connected
probes.
When using gamma
probes, the dose rate shown on the LC-display is Sv/h. When an
end window probe for α- and β- measurements or a
NaJ-scintillation probe for β-/γ-measurements is connected, the
measured value is displayed in cps. Additionally to the digital display,
the dose rate is also indicated in analog form by a logarithmic bar
graph.
The instrument is autoranging. When a measuring channel triggers a dose
rate alarm, the integrated alarm light of the "WS 05 C" gives an optical
as well as an acoustical alarm. In the standard version: orange for
"alarm threshold exceeded" and red for "error" (e. g. defective probe).
The red LED situated above the display indicates, by which measuring
channel the alarm has been triggered.
For connecting the "WS 05 C" with a computer, an RS-232-interface for
one measuring channel is available.
|
 |
|
Relay output for an
additional external alarm lamp for each measuring channel or a potential
free relay output for each measuring channel instead of the relay
output for the external alarm lamps (max. switching voltage/current:
24V, 20 mA).
Disconnectible acoustic
alarm (only triggered when an alarm threshold is exceeded and the
monitored room's door is open)
Emergency power supply
(300W) for the uninterruptible operation of the "WS 05 C" in case of
mains failure.
Integrated alarm lamp
(optical/acoustical) for the indication of the following conditions: green: "ready for operation"
orange: "alarm threshold exceeded"
red: "error"
External alarm lamps
((optical/acoustical)
Probe cable
(standard length: 1,25 m), extension up to 100 m
Special
versions upon request
|
|
Technical Data "WS 05 C" |
|
Types: |
WS 05 C-1 for the connection of one probe (1 channel)
WS 05 C-2 for the simultaneous connection of max. 2
probes (2 channels)
WS 05 C-3 for the simultaneous connection of max. 3
probes (3 channels) |
|
Probes: |
The probes are preferably connected by means of a probe cable
(up to 100 m) or
directly via the sockets of the of the "WS 05 C". |
|
Measuring size: |
H*(10) ambient dose rate equivalent for CE-probes |
Available probes:
18545 CE:
18550 CE:
18529 CE:
18509 CE:
18526 D:
NaJ-Scintillation Probe 2002:
|
Indicating range PTB-approved
measuring range
60 nSv/h - 200 µSv/h 150 nSv/h - 200 µSv/h
(γ-radiation and X-rays)
100 nSv/h - 20 mSv/h 10 µSv/h - 20 mSv/h
(γ-radiation and X-rays)
10 µSv/h - 10 Sv/h 0,5 mSv/h - 10 Sv/h
(γ-radiation and X-rays)
10 µSv/h - 1 Sv/h 50 µSv/h - 1 Sv/h
(γ-radiation and X-rays)
cps, α-, β-, γ-radiation
cps, β-, γ-radiation
PLEASE NOTE:
The PTB-approval of the probes is only valid when used together
with the dose rate meter "X5Cplus" |
|
Measuring accuracy: |
≤ +/- 10% |
|
Display: |
each measuring channel is equipped with an illuminated
LC-display |
|
Indication: |
- digital indication in Sv/h or cps, depending on the connected
probe
- analog indication by a logarithmic bar graph |
|
Alarm: |
- probe failure alarm
- dose rate alarm thresholds 1 - 4
- dose alarm thresholds 1 - 4 |
|
Alarm signals: |
- integrated optical/acoustical signal, triggered when a
connected probe signalizes a dose rate alarm (orange) and/or
error message (red)
- red LED for alarm indication when an alarm threshold is
triggered (at each measuring channel) |
|
Interface: |
RS-232 for 1 measuring channel (WS 05 C-2 / WS 05 C-3 → first
measuring channel) |
|
Dose rate alarm thresholds: |
for each measuring channel 4 dose rate alarm thresholds are free
programmable over the whole measuring range of the connected
probes |
|
Dose alarm thresholds: |
for each measuring channel 4 dose alarm thresholds are free
programmable over the whole measuring range of the connected
probes |
|
Temperature range: |
0 °C up to +40 °C for the area monitor
-30 °C up to +60 °C for the probes |
|
Humidity: |
max. 90% relative humidity |
|
Power supply: |
90-260V AC 200 mA max. 50/60 Hz |
|
Housing: |
plastic, high impact |
|
Dimensions: |
260 mm x 230 (455) mm x 150 mm |
|
Weight: |
approx. 2,8 kgs |
REMARK:
The electronics used for each single measuring channel is taken from the
PTB-approved dose rate meter "X 5 C plus".
|
Per
richiedere un preventivo per l'acquisto di
un misuratore di radioattivitą (dosimetro),
compila il form |
|