the extension of the basic
unit's measuring range
-
for
measurements at "hard-to-get-to" places
-
for
measuring high dose rates from a safe
distance
-
automatic probe identification by the basic unit
-
connectible to the basic unit either directly or by using a probe
cable up to a max. length of 100 m (standard length: 1,25 m)
-
basic
unit automatically takes over the probe specific calibration factor
-
display of H*(10) for a connected CE-probe
and display of Hx for a connected DE-probe
-
for
all connected PTB-approved H*(10)-probes, additional display of the
PTB approval symbol
-
pulse or end-window probes for the
detection of alpha, beta and gamma contaminations (cps)
-
basic
unit automatically displays alpha, beta or gamma sensitivity of the
connected probe
-
immersion counter tube ZP 1083 D for
measurements in liquids
-
underwater measurements up to a depth of 30 m by using the
pressurized water protective housing
-
Scintillation Probe 2002 with high
sensitivity (cps)
-
Contamination Probe ABG-170 with large
detector surface (plastic scintillator) for the detection of alpha,
beta and gamma contaminations (cps)
-
Telescope Probe DE for measuring
high dose rates from a safe distance
-
"CE"-Telescope to be combined with
various probe heads of the GRAETZ probe programme (according to the
required mesuring task)
-
temperature range for GM-probes: -30 °C up to +60 °C
-
temperature range for the Scintillation Probe: -20 °C up to +50 °C
(temperature change max. 10 °C/hour)
|
Type |
Energy range |
Measuring range |
Measuring size |
PTB-Approval No. |
Dimensions
Weight |
|
|
18509 CE |
55 keV - 1,3 MeV |
50 µSv/h - 1 Sv/h |
H*(10) |
23.71/05.01 |
length 110 mm, diameter 40 mm
150 g |
 |
|
18529 CE |
70 keV - 3 MeV |
0,5 mSv/h - 10 Sv/h |
H*(10) |
23.71/05.02 |
length 110 mm, diameter 40 mm
150 g |
 |
|
18545 CE |
40 keV - 1,3 MeV |
150 nSv/h - 200 µSv/h |
H*(10) |
23.71/05.03 |
length 345 mm, diameter 25/40 mm
380 g |
 |
|
18550 CE |
40 keV - 1,3 MeV |
10 µSv/h - 20 mSv/h |
H*(10) |
23.71/05.04 |
length 110 mm, diameter 40 mm
150 g |
 |
|
Type |
Energy range |
Measuring range |
Dimensions
Weight |
|
|
18509 DE |
48 keV - 2 MeV |
100 µSv/h - 1 Sv/h |
length 110 mm, diameter 40 mm
150 g |
 |
|
18529 DE |
80 keV - 2 MeV |
500 µSv/h - 10 Sv/h |
length 110 mm, diameter 40 mm
150 g |
 |
|
18545 DE |
40 keV - 1,3 MeV |
100 nSv/h - 200 µSv/h |
length 345 mm, diameter 25/40 mm
380 g |
 |
|
18550 DE |
48 keV - 1,3 MeV |
5 µSv/h - 20 mSv/h |
length 110 mm, diameter 40 mm
150 g |
 |
-
probes for α-, β- and γ-contaminations
-
high
sensitive scintillation probe for the detection of β-/γ-radiation
-
high
sensitive contamination probe for the detection of α-, β- and
γ-contaminations
-
immersion counter tube ZP 1083 D for
measurements in liquids
-
automatic probe identification by the basic unit
-
connectible to the basic unit either directly or by using a probe
cable up to a max. length of 100 m (standard length: 1,25 m)
-
basic
unit automatically takes over the probe specific calibration factor
-
indication range at the basic unit 0-20 Kcps
-
difference between measurement with pulse probes and gamma probes:
- basic unit measures pulse rates instead of dose rates,
- summation of counts instead of dose measurement,
- one pulse respectively pulse rate alarm can be adjusted at the
basic unit instead of four dose and dose rate alarm thresholds
-
temperature range for GM-probes: -30 °C up to +60 °C
-
temperature range for the Scintillation Probe: -20 °C up to +50 °C
(temperature change max. 10 °C/hour)
|
Type |
Kind of radiation |
Detector |
Background counts* (counts/min) |
Dimensions
Weight |
|
|
18526 D |
α, β, γ |
GM-tube
effective surface 6,1 cm2
|
25 |
length 110 mm, diameter 40 mm
150 g |
 |
|
ZP 1083 D |
β, γ |
GM-tube
effective length 64 mm |
15 |
length 210 mm, diameter 50 mm
350 g |
 |
* at 0,1 µSv/h
-
high
sensitivity
-
large
detector surface
-
application fields: nuclear medicine, environmental monitoring,
security controls
(police, customs, border authorities, factory inspectorates)
-
by
means of a probe cable connectible to the GRAETZ dose rate meter
X 5 C
plus and predecessor models X 5 C, X 5 DE and X 50 DE; probe
identification as pulse probe
-
by
means of a probe cable connectible to the contamination monitor
CoMo
170
-
power
supply from the basic unit
|
 |
|
Technical Data |
|
Type of radiation: |
β, γ |
|
Energy range: |
γ: 25 keV - 2 MeV
β: > 100 keV |
|
Detector: |
(70 x 70 x 13) mm
NaJ(TI)-scintillator with photo multiplier |
|
Reference orientation: |
vertical onto the square surface |
|
Position of the detector: |
middle of the detector cap |
|
Reference point: |
middle of the square surface |
Background count rate:
(at 0,1 µSv/h) |
approx. 100 cps |
|
Range of relative humidity: |
0 up to 90%, no influence |
|
Range of outside air pressure: |
600 up to 1300 hPa, influence negligible |
|
Indication ranges: |
0-200 cps
200-2000 cps
2-20 kcps
(20 kcps correspond to approx. 13 µSv/h referred
to Cs 137) |
|
Temperature range: |
operation: -20 °C up to +50 °C
storage: -30 °C up to +60 °C
(NOTE: max. temperature change 10°C/h)
|
|
Overload capacity: |
up to approx. 3 mSv/h (referred to Cs 137) |
|
After effect: |
after exposure to radiation of 3 mSv/h
error after 20 minutes is approx. 5% |
|
Voltage supply: |
+5V from the basic unit |
|
Housing: |
aluminium, partly anodically oxidized
protection class IP 65 (hose-proof) |
|
Dimensions / Weight: |
(80 x 85 x 35) mm with 200 mm handle (diameter 33 mm)
530 g |
-
thin-layer plastic scintillation detector with ZnS-coating,
integrated photo multiplier and detector electronics
-
high
sensitivity
-
detection of α-, β- and γ-contaminations
-
detector surface 170 cm²
-
by
using a probe cable connectible as pulse probe to the GRAETZ dose
rate meter
X 5 C
plus
-
display: cps or pulse summation with integration time
-
power
supply from the basic unit X 5 C plus
-
acoustic single-pulse indication at the X 5 C plus
|
 |
|
Technical Data |
|
Type of radiation: |
α, β and γ |
|
Indication range: |
0 - 20 Kcps |
|
Background counts: |
15 - 25 cps |
|
Active detector surface:
|
170 cm² |
|
Alarm threshold:
|
cps and pulse summation |
|
Dimensions:
|
approx. (390x125x75) mm, handle included |
|
Weight:
|
approx. 790 g |
-
telescopic probe for measuring high dose rates from a safe distance,
dose rate measuring range up to 10 Sv/h
-
available for the measuring sizes:
1) Photon dose equivalent rate Hx
alternatively
2) ambient dose equivalent rate H*(10)
-
for
measurements at "hard-to-get-to"-places
-
simple adaption to the basic units (X 5 C
plus, X 5 C, X 5 DE or X 50 DE) by snap-in
technique
-
reliable connection between Telescope Probe DE and basic unit
without plug or cable, only by contacts
-
telescope made of stainless steel, smoothly extendable up to 4 m
-
untwistable connection between the head of the detector probe and
the electronics
-
for
measuring low and high dose rates the Telescope Probe DE is equipped
with two different GM-tubes
-
automatic range change
-
indication of the dose rate at the user's location selectable upon
keystroke
-
no
need of separate batteries for the Telescope Probe DE, voltage
supply from the basic unit
-
especially approved for the use at German Fire Brigades
|
 |
|
Technical Data |
|
Energy range: |
low dose rate 60 keV - 1,3 MeVhigh dose
rate 60 keV - 3 MeV |
|
Indication range: |
0 nSv/h - 10 Sv/h |
|
Measuring range: |
1,5 µSv/h - 10 Sv/h |
|
Dose rate alarm thresholds: |
like basic unit |
|
Power supply: |
from the basic unit |
|
Temperature range: |
operation: -30 °C up to +60 °Cstorage:
-40 °C up to +70 °C |
|
German Fire Brigades Approval No.: |
DL/FW/GSF 06 89 11 |
|
Dimensions: |
920 mm / extended: 4000 mm |
|
Weight: |
2600 g |
-
for
measurements at "hard-to-get-to" places
-
for
measuring high dose rates from a safe distance
-
according to the required measuring task the user may operate with
various probes of the GRAETZ programme
-
untwistable connection between the probe plug connector and the
electronics
-
telescope made of stainless steel, smoothly extendable up to 4 m
-
simple and safe adaption to the PTB-approved basic unit
X 5 C
plus without needing a plug or cable, only via the
contacts by snap-in technique
(probe and basic unit DO NOT BELONG to the CE-Telescope; please
order separately)
-
indication of the dose rate at the user's location, selectable upon
keystroke
-
power
supply from the basic unit
X 5 C
plus
-
Weight: approx. 2600 g
|
 |
|
Probe Type |
Radiation |
Energy Range |
Measuring Range |
Measuring Size |
PTB-Approval |
|
18509 CE |
γ |
55 keV - 1,3 MeV |
50 µSv/h - 1 Sv/h |
H*(10) |
23.71/05.01 |
|
18529 CE |
γ |
70 keV - 3 MeV |
0,5 mSv/h - 10 Sv/h |
H*(10) |
23.71/05.02 |
|
18545 CE |
γ |
40 keV - 1,3 MeV |
150 nSv/h - 200 µSv/h |
H*(10) |
23.71/05.03 |
|
18550 CE |
γ |
40 keV - 1,3 MeV |
0 µSv/h - 20 mSv/h |
H*(10) |
23.71/05.04 |
|
18526 D |
α, β, γ |
|
0-20 Kcps background: 25 counts/min
|
|
|
|
X 5 C plus |
γ |
40 keV - 1,3 MeV |
1,0 µSv/h - 20 mSv/s |
H*(10) |
23.51/04.01 |
|
Per
richiedere un preventivo per l'acquisto di